<5. Parameter Setting>

5-13

IM 01C25T04-01EN

h. Bar Graph

ON/OFF setting of Bar Graph in upper field of

integral indicator can be performed by

Bar Graph

parameter.

• Procedure to call up the display

DTM, EDD

Local Display → Bar Graph →

→ Enable

Enable to indicate bar graph

→ Disable

Disable to indicate bar graph

i. Display Cycle

Displaying period can be set from

Display Cycle

parameter.

• Procedure to call up the display

DTM, EDD

Local Display → Display Cycle →

Recommend

Display cycle: 2.8 s

Update period of process value is 400 ms

→ 1-10

Display cycle:

400 ms × (Number selected)