<8. Diagnostic Information>

8-9

IM 01C25T04-01EN

ALARM

Display

Cause of

Alarm

SENSOR Transducer block

NAMUR

NE107

Category

Pressure Value

Primary Value

Calculated Value

Static Pressure Value Capsule Temp

Amplifier Temperature

Value

Status

Value

Status

Value

Status

Value

Status

AL.40

PB O/S

Physical block

is O/S mode

Measured

Value

BAD: non

specific

Measured

Value

BAD: non

specific

Measured

Value

BAD: non

specific

Measured

Value

BAD: non

specific

---

AL.41

STB O/S

SENSOR

Transducer

block is O/S

mode

BAD: out of

service

BAD: out of

service

BAD: out of

service

BAD: out of

service

AL.43

AI O/S

AI1 function

block is O/S

mode

GOOD (NC):

OK

GOOD (NC):

OK

GOOD (NC):

OK

GOOD (NC):

OK

AL.44

AI O/S

AI2 function

block is O/S

mode

AL.45

AI O/S

AI3 function

block is O/S

mode

AL.72

TOT O/S

Totalizer

function block

is O/S mode

AL.50

P.SDEV

P.ZDEV

Trimming

range error

for differential

pressure span

Measured

Value

UNCERTAIN:

engineering

unit violation

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

C

AL.51

SP.SDEV

SP.ZDEV

Trimming

range error for

static pressure

span

GOOD (NC):

OK

UNCERTAIN:

engineering

unit violation

AL.79

LCD.

RNG

Displayed

value exceeds

limit or LCD

configuration

error

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

---

AL.90

AI SIM

Simulation

function of AI1

block is active

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

C

AL.91

AI SIM

Simulation

function of AI2

block is active

AL.92

AI SIM

Simulation

function of AI3

block is active

AL.93

AI MAN

AI1 block is

MAN mode

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

C

AL.94

AI MAN

AI2 block is

MAN mode

AL.95

AI MAN

AI3 block is

MAN mode

AL.96

STB.SIM

Simulation

function of

Sensor TB is

active

UNCERTAIN:

Simulated

Value

AL.97

PB SIM

Simulation

function of

Physical block

is active

GOOD (NC):

OK

AL.72

TOT

MAN

Totalizer

Functiion

block is MAN

mode

AL.60

STB.

CFG

SENSOR

Transducer

block

configuration

Error

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

C