2 bus power on, 3 confirmation of device information, Bus power on -2 – Yokogawa EJX115A User Manual

Page 18: Confirmation of device information -2

Advertising
background image

<4. Getting Started>

4-2

IM 01C25T04-01EN

4.2 Bus Power ON

Turn on the power of the host and the bus.
For models with the integral indicator code “E”, the

display shows all segments in the LCD and then

changes to the displays shown below sequentially.

All segments display

F0403.ai

Model name (3 s)

Communication Protocol (3 s)

Device Revision (3 s)

Ident Number (3 s)

P

SP

T

F

P

SP

T

F

P

SP

T

F

P

SP

T

F

P

SP

T

F

Either “EJX” or “EJA” is displayed on the “Model

name” display depends on the model of transmitter.

NOTE

LCD display can be set to all segments display

only by changing the following parameter.

LCD Tranceducer Block Power On Information

(POWER_ON_INFO)

1:ON(Default)

Show all segments display, Model name,

Communication Protocol, Device Revision and

Ident Number when powering on.

0:OFF

Show all segments display when powering on.

• Procedure to call up the display

DTM, EDD

Local Display → Power On Information

ON

Enable the power on information

OFF

Disable the power on information

4.3 Confirmation of Device

Information

Software Revision, Device Revision and Ident

Number can be confirmed by following procedure.

(1) Software Revision
(a) Confirmation by the name plate

Software Revision is marked on the name plate.

Refer to Figure 4.3.

(b) Confirmation by parameter

Software Revision can be confirmed from

Physical Block parameter Software Revision

(SOFTWARE_REVISION).

• Procedure to call up the display

DTM

Physical Block → Identification →

Software Revision

EDD

Identification → Device → Software

Revision

(2) Device Revision
(a) Confirmation by the name plate

Device Revision is marked on the name plate.

Refer to Figure 4.3.

(b) Confirmation by parameter

Device Revision can be confirmed from

parameter Device Revision Compatibility or

Device Revision. (DEV_REV_COMP, DEV_

REV)

• Procedure to call up the display

DTM

[Menu] → Query Device → Query

Device PB →

EDD

[Menu] → View → Device Status

→ General → Physical Block →

→ Device Revision

Compatibility

Compatibility of Device Revision

→Device Revision

Device Revision

Advertising