<5. Parameter Setting>

5-2

IM 01C25T04-01EN

5.1.3 Output Mode (Characterization Type)

The Output Mode (Characterization Type) is set

as specified in the order when the instrument is

shipped. Follow the procedure below to change

the Output Mode. Sensor Transducer Block Target

Mode need to be changed to Out of Service (O/S)

before changing Output Mode(Characterization

Type).

DTM, EDD

Easy Setup → Sensor Transducer Block

→ Target Mode

• Procedure to call up the display

DTM, EDD

Easy Setup → Sensor

Transducer Block →

Characterization Type →

→ Linear

Linear output

→ User defined

(table)

Signal Characterizer

→ Square root

Square root output

5.1.4 Output Scale (Out Scale) and Unit

Analog Input 1 Function Block Target Mode need

to be changed to Out of Service (O/S) before

changing Output Scale(Out Scale) and Output

Scale Unit.

DTM

Easy Setup → Analog Input 1 → Target

Mode

EDD

Easy Setup → Function Block 1 – Analog

Input → Target Mode

(1) Output Scale (Out Scale)

• Procedure to call up the display

The Out Scale is factory-set as specified by the

customer. To change the range, follow the steps

below.

DTM

Easy Setup → Analog Input 1 →

→ LRV

Out Scale:Lower Value

→ URV

Out Scale:Upper Value

EDD

Easy Setup → Function Block 1 – Analog

Input → Output Scale

→ LRV

Lower Value

→ URV

Upper Value

(2) Out Scale Unit
The unit parameter is set at the factory before

shipment if specified at the time of order. Follow the

procedure below to change the unit parameter.

• Procedure to call up the display

DTM

Easy Setup → Analog Input 1 → Out

Scale:Unit

EDD

Easy Setup → Function Block 1 – Analog

Input → Output Scale → Unit

5.1.5 Damping time constant (Filter Time

Const)

The damping time constant is set at the factory

before shipment if specified at the time of order with

option code /CD. Follow the procedure below to

change the damping time constant. Target Mode of

Analog Input 1 Function Block need to be changed

to Out of Service (O/S).

DTM

Easy Setup → Analog Input 1 → Target

Mode

EDD

Easy Setup → Function Block 1 – Analog

Input 1 → Target Mode

• Procedure to call up the display

DTM

Easy Setup → Analog Input 1 → Filter

Time Const

EDD

Easy Setup → Function Block 1 – Analog

Input 1 → Filter Time Const

5.2 Calibration

The transmitter is factory characterized, Factory

characterization is the process of comparing a

known pressure input with the output of each

transmitter sensor module over the entire pressure

and temperature operating range. During the

characterization process, this comparison

information is stored in the transmitter EEPROM. In

operation, the transmitter uses this factory-stored

curve to produce process variable output (PV), in

engineering units, dependent on the pressure input.
The calibration procedure allows you to adjust

for local conditions, changing how the transmitter

calculates process variables. There are two ways

to calibrate: zero point adjustment typically used to

compensate for mounting position effects or zero

shifts caused by static pressure. A full calibration is

a two-point process, in which two accurate end-

point pressures are applied (equal to or greater

than the range value), and all output is linearized

between them.