1 pressure calibration, Pressure calibration -3 – Yokogawa EJX115A User Manual

Page 23

Advertising
background image

<5. Parameter Setting>

5-3

IM 01C25T04-01EN

5.2.1 Pressure Calibration

(1) Zero Point Adjustment
Zero Point Adjustment can be performed in several

ways. Choose the optimum method in accordance

with the circumstances specified to the application

employed.

a. Auto zero adjustment (Lower Calibration

Point)

Lower Calibration Point carries out the zero

adjustment and automatically sets the applied “0”

input values to the output value of “0”, keeping the

span constant. Use this setting when the LRV is

known to be “0”. Calibration
Target Mode need to be changed to Out of Service

(O/S) before zeroing by Lower Calibration Point.

DTM

Zero Point Adjustment → Target Mode

EDD

[Menu] → Device → Zero Point

Adjustment → Target Mode

• Procedure to call up the display

DTM

Zero Point Adjustment

EDD

[Menu] → Device → Zero Point

Adjustment →

→ Set Zero

(Method)

Adjust the lower point

b. Manual zero adjustment (Lower Calibration

Deviation)

This zero adjustment calibrates the transmitter

output corresponding to the reference pressure.

For Tank level measurement for example, enter the

actual tank level data which have determined by

glass gauge or other level measuring instrument.

• Procedure to call up the display

DTM

Calibration → Pressure

Calibration → Calibration →

EDD

[Menu] → Device → Sensor

Calibration → Pressure

Calibration → Calibration →

→ Lower Calibration

Deviation

Adjust the lower point

F0501.ai

25.00 kPa

0.00 kPa

Actual level

13.50 kPa

DPharp span: 0 to 25.00 kPa

Actual level: 13.50 kPa

Transmitter output: 13.83 kPa

DPharp

c. Zero Point Adjustment by using adjustment

Screw

Follow the procedure below to enable or inhibit zero

point adjustment by means of the external zero-

point adjustment screw on the transmitter. Local

Operation setting need to be changed to “Enabled”

before adjustment.

• Procedure to call up the display

DTM

Physical Block → Configuration →

Local Operation →

EDD

Identification → Device → Local

Operation →

→ Enabled

Enable the external zero point

adjustment

→ Disenabled Disenable the external zero point

adjustment

(2) Full Adjustment
The full adjustment is a two-point adjustment,

and the lower point adjustment should always be

performed before the upper point adjustment in

order to maintain the pitch between the zero and

100% points within the calibration range.
In the manual method, the reference pressure

should also be applied to the transmitter at both

the lower and upper points. Without the reference

pressure, Lower calibration deviation and Upper

calibration deviation may not represent the correct

value for each adjustment points.

Advertising