3 alarm mask function, 3 simulation function, Alarm mask function -2 – Yokogawa EJX115A User Manual

Page 43: Simulation function -2

Advertising
background image

<7. In-Process Operation>

7-2

IM 01C25T04-01EN

7.2.3 Alarm Mask Function

Table 7.1 shows the parameters related

to the masking of alarms. Users can mask

DIAGNOSIS_EXTENSION by setting the Diagnosis

Mask Extension RW (DIAGNOSIS_MASK_

EXTENSION_RW) parameter. DIAGNOSIS_

EXTENSION displays the results of masking.
For details of DIAGNOSIS_EXTENSION, see Table

8.3.
The DIAGNOSIS_MASK and DIAGNOSIS_MASK_

EXTENSION are a read-only parameter to display

bits which are supported by the device.

• Procedure to call up the display

DTM

[Menu] → Device → Configuration →

Query Device → Query Device PB →

EDD

[Menu] → Device → Query Device →

Physical Block → Index60 - →

Diagnosis Mask Extension RW

Table 7.1

Mask parameter of alarm in the physical block

Index

Parameters

Description

Valid Range

Initial Value

31

DIAGNOSIS_MASK

Supported bit of DIAGNOSIS

Read Only

0x00, 0x98, 0x0f,

0x80

32

DIAGNOSIS_MASK_

EXTENSION

Supported bit of DIAGNOSIS_EXTENSION

Read Only

0xff, 0x3c, 0x37,

0xf2, 0x00, 0x00

62

DIAGNOSIS_MASK_

EXTENSION_RW

Set masked DIAGNOSIS_EXTENSION.

When DIAGNOSIS_MASK_EXTENSION_RW-bit is ON and its

alarm or warning occurs, its DIAGNOSIS_EXTENSION-bit is ON.

0: Off,

1: On

0xff, 0x3c, 0x33,

0xf0, 0x00, 0x00

7.3 Simulation Function

It is possible to conduct testing for the downstream

function blocks or alarm processes. Following

description is how to use and how to set parameters

of this function.
A SIMULATE_ENABLE switch is mounted in

the transmitter amplifier. This is to prevent the

accidental operation of this function. When this is

switched on, simulation is enabled. (See Figure7.2.)

To initiate the same action from a remote terminal,

if REMOTE LOOP TEST SWITCH is written to the

SIM_ENABLE_MSG parameter (index 51) of the

physical block, the resulting action is the same as is

taken when the above switch is on. In the simulation

status, an alarm is generated from the physical

block, and other device alarms will be masked;

for this reason the simulation must be disabled

immediately after using this function.

F0702.ai

Amplifier Assembly

1

2

O

N

"OFF" during operation

Write lock switch

SIMULATE_ENABLE

Figure 7.2

SIMULATE_ENABLE Switch Position

Advertising