2 status of each parameter in failure mode, Status of each parameter in failure mode -5 – Yokogawa EJX115A User Manual

Page 51

Advertising
background image

<8. Diagnostic Information>

8-5

IM 01C25T04-01EN

8.2 Status of Each Parameter in Failure Mode

Following tables summarize the value of parameters when LCD display indicates an Alarm.

Table 8.4

Action of each parameters in failure mode related Sensor Transducer block (Condensed status)

ALARM

Display

Cause of

Alarm

SENSOR Transducer block

NAMUR

NE107

Category

Pressure Value

Primary Value

Calculated Value

Static Pressure Value Capsule Temp

Amplifier Temperature

Value

Status

Value

Status

Value

Status

Value

Status

AL.01

CAP.

ERR

Pressure

Sensor

problem

Measured

Value

BAD:

Maintenance

Alarm, more

diagnosis

available

Measured

Value

BAD:

Maintenance

Alarm, more

diagnosis

available

Measured

Value

BAD:

Maintenance

Alarm, more

diagnosis

available

Measured

Value

BAD:

Maintenance

Alarm, more

diagnosis

available

F

Capsule

Temperature

Sensor

problem
Capsule

memory

problem

Hold

Value

Hold

Value

AL.02

AMP.

ERR

Amplifier

Temperature

Sensor

problem

Measured

Value

Measured

Value

Amplifier

memory

problem

Hold

Value

Hold

Value

Hold

Value

Amplifier

problem

AL.10

PRESS

Input Pressure

is outside

measurement

range limit of

capsule

Measured

Value

UNCERTAIN:

Process

related, no

maintenance

Measured

Value

UNCERTAIN:

Process

related, no

maintenance

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

S

AL.11

ST.PRSS

Static Pressure

exceeds limit

AL.12

CAP.

TMP

Capsule

Temerature is

out of range

UNCERTAIN:

Process

related, no

maintenance

AL.13

AMP.

TMP

Amplifier

Temperature is

out of range

GOOD (NC):

OK

GOOD (NC):

OK

GOOD (NC):

OK

UNCERTAIN:

Process

related, no

maintenance

AL.30

HI.HI

LO.LO

Hi_Hi or Lo_Lo

Alarm occurs

in AI1 Function

block

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

Measured

Value

GOOD (NC):

OK

---

AL.31

HI.HI

LO.LO

Hi_Hi or Lo_Lo

Alarm occurs

in AI2 Function

block

AL.32

HI.HI

LO.LO

Hi_Hi or Lo_Lo

Alarm occurs

in AI3 Function

block

AL.33

HI.HI

LO.LO

Hi_Hi or Lo_Lo

Alarm occurs

in Totalizer

Function block

Advertising