4. INSTALLATION

Horizontal pipe mounting

Vertical pipe mounting

Transmitter
mounting bolt

U-bolt nut

Mounting bracket

50mm (2-inch) pipe

U-bolt

F0403.EPS

Figure 4.3.1 Transmitter Mounting

IMPORTANT

The transmitter should be installed at least 700
mm (when the model code of the material of the
wetted part is H, at least 1300 mm) below the
process connection to ensure a positive head
pressure of fill fluid. If it can not be installed at
least 700 mm below the process connection,
please use the equation below:

h=

ϫ

7.5

ϫ

10

–3

[mm]

(P–P0)

ϫ

dHg

ds

h:

Vertical height between the process
connection and the transmitter (mm)
h

0: Install the transmitter at least h (mm)

below the process connection

h>0: Install the transmitter at most h (mm)

above the process connection

P:

Pressure in the tank (Pa abs)

P0: Minimum working pressure limit of the

transmitter (Pa abs)

If the ambient temperature range is
–10 to 50

°

C.

5254 (Wetted parts material code S)
6980 (Wetted parts material code T)
13019 (Wetted parts material code H)
6980 (Wetted parts material code U)

ds: Specific gravity of fill fluid (at 25

°

C), refer

to GS 1C22J3-E.

dHg:Specific gravity of the Mercury 13.6 (at

25

°

C)

F0404.EPS

P

h

0

(+)

(–)

Figure 4.3.2 Example of Installation to Tank (Caution on

Installation)

Advertising