Yokogawa EJA438 User Manual

Page 58

Advertising
background image

IM 1C22J1-01E

9-6

9. GENERAL SPECIFICATIONS

Item

Description

Code

T0907.EPS

Gold-plated diaphragm

T

Teflon film

(Note 2)

With FEP film and fluorinated oil
Working range: 20 to 150

°

C, 0 to 2 MPa {0 to 20 kgf/cm

2

} (Not usable under vacuum)

R

Operating temperature
correction

(Note 3)

Adjusting range: 80 to 300

°

C

V

Capillary without PVC
sheaths

When ambient temperature exceeds 100

°

C, or use of PVC is prohibited

C1

Failure alarm down-scale

(Note 4)

Output status at CPU failure and hardware error: –5%, 3.2 mA DC or less.
When combining with Optional code F1, output signal is –2.5%, 3.6 mA DC or less.

E1

A1

Stainless steel amplifier
housing

(Note 5)

Gold-plate

Amplifier housing material: JIS SCS14A stainless steel (equivalent to JIS SUS316 cast
stainless steel or ASTM CF-8M)

Update time: 0.125 sec or less, see GS for the response time

F1

Fast response

Oil-prohibited use

K5

Degrease cleansing and dehydrating treatment

Oil-prohibited use with
dehydrating treatment

Y

Sealant (liquided silicone rubber) is coated on surfaces of JIS SUS630 nuts used for cover
flange mounting.

Sealing treatment to JIS
SUS630 nuts

Painting

Lightning protector

K1

Degrease cleansing treatment

No serration

(Note 1)

Q

No serration work on the flange gasket surface (for ANSI flange only)

A

Transmitter power supply voltage: 10.5 to 32 V DC (9 to 32 V DC for Fieldbus
communication type) Allowable current: Max. 6000 A (1x40

µ

s),

Repeating 1000 A (1x40

µ

s) 100 times

P

X1

Amplifier cover only

D1

D3

P calibration (psi unit)

bar calibration (bar unit)

(See Table for Span and

Range Limits)

Calibration units

D4

M calibration (kgf/cm

2

unit)

Epoxy resin-baked coating

Color change

Coating change

Mill Certificate

Process flange, Block

Process flange, Block, Pipe, Base

( Test Pressure )

2 MPa {20 kgf/cm

2

}

3 MPa {30 kgf/cm

2

}

3 MPa {30 kgf/cm

2

}

3 MPa {30 kgf/cm

2

}

3 MPa {29.8 kgf/cm

2

}

3 MPa {30 kgf/cm

2

}

3 MPa {30 kgf/cm

2

}

2 MPa {20 kgf/cm

2

}

5 MPa {50 kgf/cm

2

}

10 MPa {100 kgf/cm

2

}

7 MPa {70 kgf/cm

2

}

14 MPa {140 kgf/cm

2

}

3 MPa {29.8 kgf/cm

2

}

7.7 MPa {77 kgf/cm

2

}

7 MPa {70 kgf/cm

2

}

14 MPa {140 kgf/cm

2

}

( Flange rating )

JIS 10K

JIS 20K

JIS 40K

JIS 63K

ANSI/JPI Class 150

ANSI/JPI Class 300

ANSI/JPI Class 600

JIS 10K

JIS 20K

JIS 40K

JIS 40K

JIS 63K

ANSI/JPI Class 150

ANSI/JPI Class 300

ANSI/JPI Class 300

ANSI/JPI Class 600

(Applicable model)

EJA438W/EJA438N

EJA438W

EJA438W

EJA438W

EJA438N

EJA438W

EJA438W

EJA438N

EJA438W

M05

M06

T41

T42

T43

T45

T46

T47

T49

T31

T32

T33

T34

T35

T36

T37

T38

T39

For model EJA438W

For model EJA438N

For A-capsule

For B-capsule

Nitrogen(N

2

) Gas

Retention time:

10 minutes

EJA438W/EJA438N

EJA438W/EJA438N

EJA438W/EJA438N

Pressure test/Leak test
Certificate

Note 1: This item cannot be applied to Model EJA438W wetted parts material code H (Hastelloy C), T

(Tantalum), or U (Titanium). (In case for code H, T or U, serration work on the flange gasket
surface is not possible)

Note 2: Teflon film can only be specified for model EJA438W.
Note 3: Specify the process operating temperature for zero correction.

Example: Zero correction by process temperature 90

°

C.

Note 4: The hardware error indicates faulty amplifier or capsule. Standard output status (without /C1) is

up-scale of 110%, 21.6 mA DC or more.

Note 5: Applicable only for electrical connection code ‘2’, ‘3’ or ‘4’. Not applicable for optional code P

and X1. Not applicable for optional code JF1 and JS1.

Advertising