5 self-diagnostics, 1 checking for problems, Example 2: setting entry errors – Yokogawa EJA438 User Manual

Page 44: Example 3: checking the history of the errors

Advertising
background image

IM 1C22J1-01E

7-15

7. BRAIN TERMINAL BT200 OPERATION

7.5 Self-Diagnostics

7.5.1 Checking for Problems

(1) Identifying Problems with BT200

The following four areas can be checked.
(a) Whether connections are good.
(b) Whether BT200 was properly operated.
(c) Whether settings were properly entered.
(d) History of the errors.
See examples below.

• Example 1: Connection errors

ESC

communication error

Press the key.

When the panel shown on the left

appears, press the key.

Since communications will be

unsuccessful if there is a problem

in the connection to the BT200, the

display at the left will appear.

Recheck the connection.

Press the (OK) key.

––WELCOME––

BRAIN TERMINAL
ID: BT200

check connection
push ENTER key

UTIL

FEED

• Example 2: Setting entry errors

F0736.EPS

The initial data panel shows the

result of current transmitter

diagnostics.

DATA

DIAG

PRNT

ESC

PARAM
C20:PRESS UNIT
kPa
C21:LOW RANGE
600 kPa
C22:HIGH RANGE
600 kPa

FEED

PRNT

ESC

DIAG
C60:SELF CHECK
ERROR
< ERROR >
< ILLEGAL LRV >

Press the (DIAG) key in the

parameter panel to go to the

diagnostics panel

(C60: SELF CHECK).

An error message is displayed

when an error occurs in the

diagnostics panel.

OK

PARAM
01:MODEL
EJA438W-DA
02:DAG NO.
YOKOGAWA
03:SELF CHECK
ERROR

• Example 3: Checking the history of the errors

Connect the BT200 to the

transmitter, and call item “P.”

The history of up to four errors can be stored. When the 5th

error has occurred, it is stored in “P10”. The error stored in

“P13” will be deleted, and then, the error in “P12” will be

copied to “P13”. In this sequence, the history of the most

previously occurred error will be removed from memory.

“GOOD” will be displayed if there was no previous error.

F0737.EPS

HOME

SET

ADJ

ESC

MENU
J:ADJUST
K:TEST
M:MEMO
P:RECORD

DATA

DIAG

PRNT

ESC

PARAM
P10:ERROR REC 1
ERROR
P11:ERROR REC 2
ERROR
P12:ERROR REC 3
GOOD

For the details of the messages listed below, see Table

8.5.1 Error Message Summary.

ESC

SET
P10:ERROR REC 1
ERROR
< ERROR >
< ILLEGAL LRV >
< ILLEGAL HRV >

<(a) SETUP PANEL>

Note 1: Press the key twice in the setting panel

(panel 1) to clear all error message (P10 to P13)

information.

Note 2: After two hours from when an error occurs, the error

message of that error will be recorded. Therefore,

if you switch off the transmitter within two hours from

when the error occurs, there is no history of that

error stored in the transmitter, and this function is

meaningless.

CAP MODULE FAULT

AMP MODULE FAULT

OUT OF RANGE

OUT OF SP RANGE

OVER TEMP (CAP)

OVER TEMP (AMP)

OVER OUTPUT

OVER DISPLAY

ILLEGAL LRV

ILLEGAL HRV

ILLEGAL SPAN

ZERO ADJ OVER

P10: “ERROR REC 1” displays the last error.

P11: “ERROR REC 2” displays the error one time before

the last error occurred.

P12: “ERROR REC 3” displays the error two times before

the last error occurred.

P13: “ERROR REC 4” displays the error three times before

the last error occurred.

Select P10: ERROR REC1 and

press the key to display

the error message.

Advertising