Yokogawa RAKD User Manual

Page 37

Advertising
background image

IM 01R01B30-00E-E 5th edition January 01, 2013 -00

<7. TECHNICAL DATA>

7-7

All Rights Reserved. Copyright © 2003, Rota Yokogawa

LIMIT SWITCHES IN STANDARD VERSION

(option /K1 to /K3)

Type

:

Inductive proximity switch SC2-NO

acc. DIN EN 60947-5-6

Nominal voltage :

8V DC

Output signal :

≤ 1 mA or ≥ 3 mA

Hysteresis

:

< 0.5mm

LIMIT SWITCHES IN FAIL SAFE VERSION

(option /K6 to /K10)

Type

:

Inductive proximity switch

SJ2-SN; SJ2-S1N acc. DIN EN 60947-5-6

Nominal voltage :

8V DC

Output signal :

≤ 1 mA or ≥ 3 mA

Hysteresis

:

< 0.5mm

HYSTERESIS OF LIMIT SWITCHES

Min-contact / Max-contact :
- pointer movement : ≈ 0.8 mm
- float movement :

≈ 0.8 mm

Minimum distance between 2 contacts

:

≈ 8 mm

ELECTRICAL CONNECTION (option /K1 to /K10) :

Type

:

Quickon

Cable diameter : 4 – 6 mm
Maximum cross section of core

: Ø 0.34 to 0.75 mm

2

POWER SUPPLY FOR LIMIT SWITCHES (Option /W__)

Type

: Transmitter relay

acc. DIN EN 50227 (NAMUR)

- KFA6-SR2-Ex1-W (230 V AC)

- KFA5-SR2-Ex1-W (115 V AC)

- KFD2-SR2-Ex1-W (24 V DC)

- KHA6-SH-Ex1 (115/230 V AC),

Fail Safe, only one channel

- KFD2-SH-Ex1 (24 V DC),

Fail Safe, only one channel

Power supply :

- 230 V AC ± 10%, 45-65Hz

- 115 V AC ± 10%, 45-65Hz

- 24 V DC ± 25%

Relay output

:

1 or 2 potential-free change over

contact(s)

Switching capacity

: max. 250V AC, max. 2 A

Note :
If Fail-Safe limit switch option /K6 or /K7 is ordered, for pow-
er supply option /W2E or /W4E must be selected.
If Fail-Safe limit switch option /K8, /K9 or /K10 is ordered,
for power supply option /W2F or /W4F must be selected.

SWITCHING LEVELS FOR LIMIT SWITCHES

Table 7-1 Min, Max and Min-Max-contact in standard version

Option /K1 Option /K2 Option /K3

Function Pointer

Signal

Signal

Signal

SC2-N0

SC2-N0

SC2-N0

MAX

above LV

below LV

----
----

1 mA
3 mA

1 mA
3 mA

Function Pointer

Signal

Signal

Signal

SC2-N0

SC2-N0

SC2-N0

MIN

above LV

below LV

3 mA
1 mA

----
----

3 mA
1 mA

Note: LV = Limit value

Table 7-2 Min, Max and Min-Max-contact in fail-safe version

Option /K6 Option /K7 Option /K8

Function Pointer

Signal

Signal

Signal

SJ2-SN

SJ2-SN

SJ2-SN

MAX

above LV

below LV

Fail Safe

----
----
----

1 mA
3 mA
1 mA

1 mA
3 mA
1 mA

Function Pointer

Signal

Signal

Signal

SJ2-SN

SJ2-SN

SJ2-SN

MIN

above LV

below LV

Fail Safe

3 mA
1 mA
1 mA

----
----
----

3 mA
1 mA
1 mA

Note: LV = Limit value

Table 7-3 Limit switch as Min-Min-contact in fail-safe version

Option /K9

Function Pointer

Signal

SJ2-S1N

MIN

above LV

below LV

Fail Safe

3 mA
1 mA
1 mA

Function Pointer

Signal

SJ2-SN

MIN

above LV

below LV

Fail Safe

3 mA
1 mA
1 mA

Note: LV = Limit value

Table 7-4 Limit switch as Max-Max-contact in fail-safe version

Option /K10

Function Pointer

Signal

SJ2-SN

MAX

above LV

below LV

Fail Safe

1 mA
3 mA
1 mA

Function Pointer

Signal

SJ2-S1N

MAX

above LV

below LV

Fail Safe

1 mA
3 mA
1 mA

Note: LV = Limit value