2 setting parameters, Setting parameters, Burst mode – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 20

Advertising
background image

<3. Parameters>

3-7

IM 11M12D01-51E

3.2 Setting Parameters

3.2.1 Burst Mode

The AV550G continuously sends the data stored in it when burst mode is set “On”. Either one of

measured value, % output value, or 4 to 20 mA output value can be selected and sent. The data

is sent periodically at 75 ms intervals as a digital signal when the AV550G is set in burst mode.

Therefore, communication by the HART simultaneous communicator is also possible.

Setting of Burst Mode

Call up “

Burst option” display.

1.Device setup ---> 4.Detailed setup ---> 4.Output condition ---> 1.HART output --->
4.Burst option

Display

Operation

(ENTER)

Set data to be sent.

Measured value (PV)

% output value

(% range/current)

4 to 20 mA output value

(Process vars /crnt)

Call up "

Burst mode" display.

3.HART output 3.Burst mode

1

(ENTER)

Set

"

On

"

and press ENTER

(F4).

2

(SEND)

Press SEND (F2).

F4

F4

F2

F030209-1.ai

AV550G:

Burst option

PV

% range/current

Process vars/crnt

ENTER

ESC

HELP

AV550G:

Burst mode

Off

ENTER

ESC

off

on

AV550G:

HART output

1 Poll addr 0

2 Num req preams 5

3 Burst mode

On

4 Burst option

PV

HOME

SEND

HELP

********

Advertising