<4. Maintenance>

4-1

IM 11M12D01-51E

4. Maintenance

Various sensor data can be viewed on the HART Communicator.
For more details, refer to “Section 10.1 Display” in the AV550G Instruction Manual.
Call up “

Channels” display.

1. Device setup ---> 4. Detailed setup ---> 2. Channels

Table 4.1 Meaning of Data Labels in “Sensors”

1

2

F030901-1.ai

STU

1

AV550G:

Channels

1 Ch1 detailed setup

2 Ch2 detailed setup

3 Ch3 detailed setup

4 Ch4 detailed setup

5 Ch5 detailed setup

HOME

SAVE

AV550G:

Ch1 detailed setup

1 Ch1 status

2 Sensors

3 Error

4 Alarm

HOME

SAVE

3

AV550G:

Sensors

1 Cell mV -0.02 mV

2 Cell temp 748 degC

3 CJ temp 0 degC

4 TC volt 31.17 mV

5 CJ val

996.6 ohm

HOME

SAVE

HELP

VWX

2

Select “ ” (this

example shows how to display

various data from Channel1).

Select “ ”

Various data from Channel 1 is

displayed. Meanings of data

labels in “ ” displayed on the

HART communicator are shown in

the table below.

Operation

Display

Ch1 detailed setup

Senors

Senors

HART Communicator

display

Meaning of data

Cell mV
Cell temp
CJ temp
TC volt
CJ val
Heater duty

Cell emf displayed
Cell temperature is displayed
Cold junction (reference junction) temperature is displayed
Thermocouple emf is displayed
Cold junction (reference junction) resistance is displayed
Heater duty cycle ON time is displayed

T3009-1.ai