5 blow back – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 29

Advertising
background image

<3. Parameters>

3-16

IM 11M12D01-51E

3.5 Blow Back

Here we explain how to excute a Semi-Auto Blow Back using HART communications.

This requires the AV550G Blow back mode to have been set to Semi-Auto or Auto.

For further details, refer to “Section 10.5 Blow Back” in the AV550G Instruction Manual.

Blow-Back-related parameters can not be set remotely using HART communication.

Blow Back can not be canceled using HART communication.

Call up “

Blow back” display.

1. Device setup ---> 2. Diag/Service ---> 4. Blow back

NOTE

1

2

(OK)

F030501-1.ai

F4

AV550G:

ABORT

ON

Do you want to

execute blow back?

AV550G:

Diag/Service

1 Loop test

2 Calibration

3 Indication check

4 Blow back

5 Calib log

HELP SAVE HOME

Confirm that you want to execute

Blow Back

Blow Back is started using parameters

(Blow Back Time, Hold Time) set in

the AV550G

Display

Operation