Multidrop mode -8, 2 multidrop mode, Setting of multidrop mode – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 21

Advertising
background image

<3. Parameters>

3-8

IM 11M12D01-51E

3.2.2 Multidrop Mode

Field devices in multidrop mode refer to the connection of several field devices on a single

communication line. Up to 15 field devices can be connected when set in multidrop mode. To activate

multidrop communication, each field device address must be changed to a unique number in the

range 1 to 15. This change deactivates the 4 to 20mA output and changes it to 4mA.

Setting of Multidrop Mode

Call up “

Poll addr” display.

1.Device setup ---> 4.Detailed setup ---> 4.Output condition ---> 1.HART output

--->1.Poll addr

Call up

“Auto Poll” display.

HART output ---> Online Menu ---> Main Menu ---> 4.Utility ---> 1.Configure Communication

---> 1.Polling

"

1

"

(ENTER) (SEND)

Set the polling address (a

number from 1 to 15) and

press

ENTER (F4).

Then Press

SEND (F2).

F4

F2

F030210-1.ai

AV550G:

Poll addr

0

0

ESC

DEL

HELP

ENTER

Display

Operation

1

(HOME)

Return "

Online Menu" with

HOME (F3).

2

Return to "

Main Menu " with

a "

previous" key.

3

Select "

Utility" .

F3

F030211-1.ai

JLK

4

AV550G:

HART output

1 Poll addr

1

2 Num req preams 5

3 Burst mode

On

4 Burst option

PV

HOME

SEND

HELP

AV550G:

1

Device setup

Online

2 PV

3 PV URV

4 PV LRV

21.64

21

0

%

%

%

SAVE

HART Communicator

1 Offline

2 Online

3 Frequency Device

4 Utility

Display

Operation

Advertising