<3. Parameters>

3-15

IM 11M12D01-51E

3.4 Indication Check

Here we explain how to excute a Semi-Auto Indication check using HART communications. This

requires the AV550G Indication check mode to have been set to Semi-Auto or Auto.

For further details, refer to “Section 10. Indication Check” in the AV550G Instruction Manual.

Indication-Check-related parameters can not be set remotely using HART

communication.

Indication Check can not be canceled using HART communication.

Call up “

Indication check” display.

1. Device setup ---> 2. Diag/Service ---> 3. Indication check

NOTE

1

2

3

(OK)

F030401-1.ai

F4

STU

1

AV550G:

Select channel.

1 Ch1

2 Ch2

3 Ch3

4 Ch4

5 Ch5

ABORT ENTER

AV550G:

ABORT

ON

You select Ch1.

Do you want to

execute semi-auto

indication check?

AV550G:

Diag/Service

1 Loop test

2 Calibration

3 Indication check

4 Blow back

5 Calib log

HELP SAVE HOME

Select Channel for Indication Check

(Ch1 in this example).

Confirm that you want to execute

Indication Check

Indication Check is started using

parameters (Check Time, Hold Time)

set in the AV550G

Display

Operation