3 calibration, Calibration -14 – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 27

Advertising
background image

<3. Parameters>

3-14

IM 11M12D01-51E

3.3 Calibration

Here we explain how to excute Semi-Auto Calibration using HART communications.

This requires the AV550G Calibration mode to have been set to Semi-Auto or Auto.

For further details, refer to “Section 9. Calibration” in the AV550G Instruction Manual.

Calibration-related parameters can not be set remotely using HART communication.

Calibration can not be canceled using HART communication.

Call up “

Calibration” display.

1. Device setup ---> 2. Diag/Service ---> 3. Calibration

NOTE

1

2

3

(OK)

F030301-1.ai

F4

STU

1

AV550G:

1 Ch1

Select channel.

2 Ch2

3 Ch3

4 Ch4

5 Ch5

ABORT ENTER

AV550G:

ABORT

ON

You select Ch1.

Do you want to

execute semi-auto

calibration?

AV550G:

Diag/Service

1 Loop test

2 Calibration

3 Indication check

4 Blow back

5 Calib log

HELP SAVE HOME

Select Channel to be calibrated

(Ch1 in this example).

Confirm that you want to execute

calibration

Calibration is started using parameters

(Cal. Time, Hold Time) set in the

AV550G

Display

Operation