IM 11M12D01-51E

3.7 Status Check

You can check Channel Card Status using the HART Communicator. The meanings of these status

indications are displayed in the table below.

Call up “

Detailed setup” display.

1. Device setup ---> 4. Detailed setup

Table 3.2 HART Communicator Display and Channel Card Status

1

2

F030801-1.ai

YZ/

3

AV550G:

Detailed setup

1 Control

2 Channels

3 Channel status

4 Output condition

5 Device information

HOME

SAVE

AV550G:

Channel status

1 Ch1 sts Measuring

2 Ch2 sts Disable

3 Ch3 sts Measuring

4 Ch4 sts Error

5 Ch5 sts

None

HOME

SAVE

Select “ ”

In the display example at left, Channel 2 is Disabled,

Channel 4 is in Error status, and Channel 5 is Not

Installed.

You can scroll the screen to display status of channels

6 and later.

Display

Operation

Channel status

HART Communicator

status display

Channel card status

Card is Not Installed
Channel is disabled
Channel is in Warm Up status
Channel is in Measurement status
Channel is in Blow Back status
Channel is in Calibration status
Channel is in Indication Check status
Channel is in Error status
Process Gas alarm has occurred.

None
Disable
Warm up
Measuring
Blow back
Cal
Ind. check
Error
Proc gas

T0302-1.ai