<3. Parameters>

3-9

IM 11M12D01-51E

1. If “Polling” is set to “Never Poll” after the address is set, “Online Menu” cannot

be called up and displayed. Be sure to set “Polling” to “Digital Poll” after setting

the polling address.

2. When the same polling address is set for two or more field devices in multidrop

mode, communication with those field devices is disabled.

Example: Communication when set in the multidrop mode.

4

Select "

Configure

Communication "

.

5

Select "

Polling " .

6

x3

(ENTER)

Select "

Digital Poll"

and press ENTER (F4).

F4

F030212-1.ai

HART Communicator

Utility

1 Configure Communic

2 System Information

3 Listen for PC

4 Storage Location

5 Simulation

HART Communicator

Configure Communica

1 Polling

2 Contrast

3 Off Time

4 Ignore diagnostics

5 Delete Configs

HELP

HART Communicator

Polling

Ask Before Polling

Never Poll

Ask Before Polling

Always poll

Digital poll

ENTER

ESC

HELP

NOTE

1

2

3

(1) The HART Communicator searches for field

devices is set in multidrop mode when the HART

Communicator is turned on.

When a HART Communicator is connected to a

field device, its tag will be displayed (display 1).

(2) Select desired field device. After that, normal

communication with the selected field device is

possible. However, the communication speed is slow

in this case (display 2).

(3) To communicate with another field device, call up

display 3 , and select "Online".

(4) Display1 will appear. Repeat the above operation.

F030213-1.ai

HART Communicator

Online

1 1: ZIA-01A

2 2: ZIA-02A

rm

AV550G:

1

Device setup

Online

2 PV

3 PV URV

4 PV LRV

21.64

21

0

%

%

%

SAVE

ZIA-01A

BACK

Display

Operation

HART Communicator

1 Offline

2 Online

3 Frequency Device

4 Utility