Laccess to the address not listed or mapped, Lretry to writing, Lcharacter time-out – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 13: Lholding time of data renewal bits, Ltcp time-out

Advertising
background image

<3. Functions>

3-4

IM 11B06E01-01E

Holding register (40001 to 49999)

Data

Address

Description

Time setting

value

40001

40002

40003

40004

Setting the clock time.

For data format, refer to “5.3 Holding Register (set data)” in “GS 11B08B02-

01E GC8000 Modbus Communication” or “GS 11B03G02-02E GC1000

MarkII Modbus Communication”.

The data are kept after writing on the analyzer and PCAS.

Mapping area 41001 to

49000

Setting the input register or holding register address of analyzers.

ASGW only reads the input register on the analyzer.

ASGW only writes the holding register on the analyzer.

The holding registers are as follows.

GC8000:

Refer to “5.3 Holding registers (setting data)” in “GS 11B08B02-01E

GC8000 Modbus Communication”. However, “40001-40004: Time Setting

Request” is not available.

GC1000 MarkII:

Refer to “5.3 Holding Register (set data)” in “GS 11B03G02-02E GC1000

MarkII Modbus Communication”. However, neither “40001-40004: Time

Setting Request” nor “41DDD: Analysis value” is available.

The other address area is ignored. The holding register in PCAS and ASIU

can not be set.

Input register (30001 to 39999)

Data

Address

Description

Mapping area

31001 to 39000 Setting the input register address of analyzers.

ASGW reads the addressed data from analyzers.

l

Access to the address not listed or mapped

Access to the address that is not shown on the table or not mapped does not generate an error.
However, the value of data is uncertain.
DCS can not write to the coil or holding register of ASGW that is not defined in the table above.

However, with setting “#ADDR_CHK, OFF” after-mentioned, DCS can write to thecoil or holding

register of ASGW.

l

Retry to writing

When an error occurs in writing to the coil or holding register of analyzers and PCAS, ASGW tries

to write the specified number of times. The number of retry times can be specified on the setting

file.

l

Character time-out

Value of character time-out for communication between DCS and Modbus RTU can be set ac-

cording to communication speed. The value of mean time of each characters is set on the setting

file.

l

Holding time of Data renewal bits

The variable data bits such as Data renewal, Renewal of Calibration Coefficient and Alarm Status

Change and be held for a constant time for DCS reading. Holding time can be specified. The unit

is second.

l

TCP time-out

Value of time-out for Modbus TCP communication. The unit is second. 1 second or more is rec-

ommended.

4th Edition : Nov. 22, 2011-00

Advertising
This manual is related to the following products: