<4. Engineering>

4-13

IM 11B06E01-01E

4th Edition : Nov. 22, 2011-00

n

Setting IP address

Start the resource configurator software on PC and wait for a message “A new controller is con-

nected”.


When the message appears, select the “Setting IP Address” on “File” menu.


IP address, Subnet mask, Gateway and Time zone parameters are set on this dialogue window.
Reboot the ASIU/ASGW after the setting.