<6. OPC DA 2.0 Interface>

6-2

IM 11B06E01-01E

4th Edition : Nov. 22, 2011-00

6.5

List of OPC Data Names

The table below shows a list of ASGW/ASIU’s data names that can be accessed via STARDOM

OPC DA2.0 server. Data names are common to ASGW and ASIU.
ASGW/ASIU’s node name designated in the setting file on the STARDOM OPC server is

assigned to “Node.”

OPC data name example 1: [email protected](312) represents the address 32312 of the

Modbus data of ASGW01.

OPC data name example 2: [email protected](1000) represents the address 03000 of the

Modbus data of ASGW01.

l

COIL (00001 to 09000)

Read/write is possible. Writing data onto unmapped areas is ignored. Unmapped address data is

always read as “FALSE”.

OPC Data Name

OPC Data Type

Function

Remark

[email protected](n)

VT_BOOL

Modbus Address 00001 to 01000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 01001 to 02000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 02001 to 03000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 03001 to 04000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 04001 to 05000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 05001 to 06000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 06001 to 07000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 07001 to 08000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 08001 to 09000

n = 1 to 1000

l

DSCI (10001 to 19000)

Read only. Writing through the OPC interface does not cause an error, but nothing can be written

onto the Modbus area of ASGW/ASIU. Unmapped address data is always read as “FALSE”.

OPC Data Name

OPC Data Type

Function

Remark

[email protected](n)

VT_BOOL

Modbus Address 10001 to 11000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 11001 to 12000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 12001 to 13000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 13001 to 14000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 14001 to 15000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 15001 to 16000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 16001 to 17000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 17001 to 18000

n = 1 to 1000

[email protected](n)

VT_BOOL

Modbus Address 18001 to 19000

n = 1 to 1000