3 setting, 3 setting -6 – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 19

Advertising
background image

<4. Engineering>

4-6

IM 11B06E01-01E

4.2.3 Setting

This file consists of text characters. Editing can be done with “Notepad” software, but “MS-Excel”

software is recommended. The edited file should be saved with “CSV” as an extension.
Each line has meaning in the map file. A line beginning with character # is interpreted as a mean-

ingful line.
And a comment line does not begin with character “# “.

NOTE

Comment lines must not include the # character, otherwise ASGW may not work properly.

#SLAVE, n

This n is the Slave ID when DCS accesses to ASGW with Modbus interface and commonly

uses on the 2 communication ports. Range is 1 to 247.

#RETRY, m

This m is the number of retry times when a write error occurs. Range is 1 to 32767.

#TOUT, t

This t is character time-out in the communication between DCS and Modbus RTU. Adjust the

value depending on communication speed. Unit is ms [millisecond] and range is 10 to 5000.

#KEEP_BIT, t

This specifies the holding time of the variable data bits. This t is the time (sec) to hold.
This function can be used on Rev. 1.01.04 or later of ASGW. The default value is 5 sec, and

range is 0 to 3276.

#TCP_TIMEOUT, t

This specifies time-out value of Modbus TCP. This t is the time-out value (sec).
This function can be used on Rev. 1.01.03 or later of ASGW. The default value is 30 sec, and

range is 1 to 100.

#TSET, StnID

Writing in address 00001 of the coil in the Modbus address table carries out time setting in all

devices that have a time setting function. If StnID is set to 255 in the #TSET line, time setting

is carried out for all devices; if StnID is set to any number from 1 to 254, the setting is effec-

tive only for a specified device.

l

Specifying analyzers and PCASs

Up to 31 analyzers and PCASs can be specified to be connected.
A unique StnID number must be allocated to each analyzer and PCAS. This StnID is key data on

Modbus interface.

Format = “#STN, StnID, StnType, IP_Address”

#STN

Key word. The line start from “#STN” is recognized as one for specifying analyzers.

StnID

ID number of analyzer (1-254)

StnType

Type: GC8000/GC1000/ASIU/GCIU/PCAS

IP_Address

IP address of analyzer

Example “#STN, 3, GC1000, 192.168.0.1”

4th Edition : Nov. 22, 2011-00

Advertising
This manual is related to the following products: