Opc da 2.0 interface, 1 system configuration, 2 software required – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 36: 3 installation and settings, 4 update rate, Opc da 2.0 interface -1, System configuration -1, Software required -1, Installation and settings -1, Update rate -1

Advertising
background image

<6. OPC DA 2.0 Interface>

6-1

IM 11B06E01-01E

6.

OPC DA 2.0 Interface

This chapter explains the OPC DA 2.0 interface of ASGW and ASIU.
Adding an OPC server of STARDOM to the system allows ASGW/ASIU to have the OPC in-

terface. There is no need to add any special settings to the ASGW/ASIU application. Map-

ping Modbus data of ASIU to ASGW allows ASIU data to be read and written by accessing

ASGW through the OPC interface.

6.1

System Configuration

GC8000

ASIU

ASGW

Ethernet

OPC Server PC

GC1000 MarkII

F0601.ai

6.2

Software Required

ASGW do not require additional software to be installed for OPC. The STARDOM software listed

below is required for a PC on which an OPC server is to be installed.

Model

Item

Remark

NT781AJ

FCN/FCJ OPC Server for Windows

NT783AJ

Redundant Network Function License

for FCN/FCJ OPC Server

Required only for redundant network.

6.3

Installation and Settings

For installation and settings of the OPC server and the duplexed package, refer to their respec-

tive user’s manuals or online help.

6.4

Update rate

Data for the OPC interface is updated every second. Therefore, a delay of about one second

may occur when the OPC client reads the latest data from analyzers.

TIP

Separating OPC data from Modbus data causes hogging of the CPU. To avoid this, the OPC data update rate is

set to 1 second while the Modbus communication is updated every 200 ms.

4th Edition : Nov. 22, 2011-00

Advertising
This manual is related to the following products: