<3. Installing the Sensor and Connecting the Cable>

3-4

IM 12J5B3-01E

3.2.3

Asembling the Sensor in the Holder

For details on how to assmble the sensor in the holder or fitting, see the appropriate instruction

manual for each holder or fitting.

CAUTION

Exercise care not to contaminate or wet the tip of the sensor cable when assembling the sensor.
If you will not begin wiring the sensor cable immediately, take the protective measures necessary

to prevent the assembly from being damaged.

Sensor mounting in the case where flocs and other matter are contained in the SUM

Stopper

F07-1.ai

The stopper projecting from the membrane assembly locknut serves to prevent the membrane

from being in contact with the ground and thus being damaged. If there are flocs in the SUM, the

stopper may become clogged. Thus, assemble the sensor to the holder so that the stopper is

positioned downstream in the SUM.

3.3

Connecting the Sensor Cable

The cable of the dissolved oxygen sensor must be connected to the terminals of the dissolved

oxygen converter. For information on connections, refer to the following instruction manu-

als: DO402G Dissolved Oxygen Converter (IM 12J05D02-01E), the DO202/FLXA21 2-wire

Transmitter instruction manual (IM 12J05C01-01E/IM 12A01A02-01E) or WTB10-DO Relay

Terminal Box (IM 12J05W02-01E).

F08.ai

IE

T1

T2

SE

RE

S

13

17

21

11

12

14

15

16

13

11

12

14

15

16

(*1)

Dissolved oxygen

converter terminals for

connection to sensor

cable

Sensor cable

Jumper

*1: Liquid earth to detect membrane failure.

This function is available when using

PB350G/PB360G float type holder.

Connect only when using this function.

Figure 3.5

Connection of Sensor Cable

Now the explanation on the installation and wiring is completed. For details on operation during

startup or for regular operation after the entire measuring system is brought into operation, refer

to the DO402G Dissolved Oxygen Converter instruction manual (IM 12J05D02-01E), the DO202/

FLXA21 2-wire Transmitter instruction manual (IM 12J05C01-01E/IM 12A01A02-01E). The main-

tenance and service of the dissolved oxygen sensor is explained in Section 4 that follows.