2 inspection if failure occurs, Caution – Yokogawa DO30G Dissolved Oxygen Sensor User Manual

Page 19

Advertising
background image

<4. Maintenance>

4-2

IM 12J5B3-01E

4.2

Inspection If Failure Occurs

If the electrolyte and/or lead electrode within the sensor unit deteriorates and ultimately com-

pletes its service life, anomalies, such as the failure to make a span adjustment, will occur. This

section deals especially with the inspection and service of the dissolved oxygen sensor needed if

any anomaly is found in a detected signal. If you fail to pinpoint the cause of failure to a particular

location, follow the instructions in Section 4.2.1 and subsequent sections in sequence. One or

more of the following measures are necessary to remedy the failure.

• Replacement of electrolyte/permeable-membrane assembly
• Replacement of sensor unit

4.2.1

Inspecting the Permeable-membrane Assembly (for

Damage)

Note : When the membrane is used in the float holder, the converter membrane damage detector function can be used. This function

generates an "E4" error message if the membrane is damaged.

Check the permeable-membrane assembly for possible wrinkling or damage. Also, check that

the permeable-membrane assembly locknut is not loose. If the locknut is securely tightened and

the membrane is normal (no wrinkling or damage is found), proceed to the instruction procedure

in Section 4.2.2. If the locknut is loosen or the membrane is wrinkled or damaged, replace the

electrolyte and the permeable-membrane assembly (including the O-ring seal) as well.
[Electrolyte/permeable-membrane assembly replacement procedure]
When replacing the electrolyte and permeable-membrane assembly, you must provide a syringe

and beaker that are used to replace the electrolyte and an abrasive and gauze to polish the silver

electrodes. Provide these materials along with the spare electrolyte solution and assembly used

in the replacement.

CAUTION

Precautions in Handling the Electrolyte
The potassium-hydroxide electrolyte (KOH) is a highly alkaline solution containing potassium

hydroxide as the major constituent. Exercise extreme care to protect your body from contact

with the solution and to prevent the solution from getting into your eyes. Should your skin or eyes

become exposed to the solution, immediately wash the exposed area with clean water. If the

solution gets in your eyes, consult an eye doctor after flushing them with water. The electrolyte

is a colorless corrosive and, therefore, there is the risk of its being mishandled. Never leave any

residual solution in the syringe.

(1) After having removed the dissolved oxygen sensor from the holder, clean dirt off the wet

part.

(2) Loosen the locknut and remove the permeable-membrane assembly. Also, remove the O-

ring.

4th Edition : Apr 05,2011

Advertising