2 checking of rtd in the sensor, 3 judgment of lead electrode fault, 3 expendables, spare parts and maintenance parts – Yokogawa DO30G Dissolved Oxygen Sensor User Manual

Page 20: 1 dox8a maintenance parts (expendables) kit

Advertising
background image

<4. Maintenance>

4-3

IM 12J5B3-01E

(3) Completely drain the electrolyte within the sensor unit. Tilt the sensor tip downward and then

use the syringe to force air into the inlet for refilling the electrolyte. Position the beaker so it

will catch any of the electrolyte that drains through the gaps around the silver electrode.

Sensor unit

Place a beaker

underneath to catch the

draining electrolyte.

Silver electrode

Syringe

Inlet for refilling electrolyte

Needle (syringe needle)

Insert the syringe needle into

the inlet for refilling electrolyte

and force air inside.

F09.ai

Figure 4.1

Draining the Electrolyte

(4) Hold the sensor level and polish the silver electrode.

Apply an abrasive (alumina) to wet gauze and polish the surface of the silver electrode until

it becomes glossy. After finishing polishing, completely wipe off any abrasive that remains

on the electrode.

(5) Make sure the sealed surfaces are free from dust. Install a new O-ring.
(6) Inject 8 ml of electrolyte into the sensor body with the syringe (see Figure 4.2).

Note: Always use an electrolyte prepared by Yokogawa for exclusive use with the DO30G.

First, fill the sensor with the electrolyte (approx. 9 ml) and then remove approximately 1 ml

so that a layer of air is present inside the sensor body.

(7) Install the permeable-membrane assembly.

It is strictly not possible to allow a layer of air to form between the membrane and the sil-

ver electrode. To prevent such a layer from forming, first apply a drop of electrolyte to the

surface of the silver electrode. Lightly press the permeable-membrane assembly so that the

O-ring touches the groove on the inside. Tighten the locknut fully without letting the assem-

bly turn.

Surface of the silver electrode

Apply a drop of electrolyte when

installing the permeable-membrane

assembly.

Fill with electrolyte

and then remove

approximately 1 ml.

Syringe

Inlet for injecting electrolyte

O-ring

F10.ai

Figure 4.2

Injecting the Electrolyte

All these steps complete the inspection procedure for the permeable-membrane assembly. By

way of precaution, make sure the membrane is free from wrinkling. Note that a stable detection

signal is not possible right after refilling with the electrolyte. Make a zero or span adjustment ap-

proximately a half a day or longer after refilling. It may indicate a minus value for about 10 min-

utes zero adjustment.