Prohibited actions, How to remove protective cap, Caution – Yokogawa DO30G Dissolved Oxygen Sensor User Manual

Page 24

Advertising
background image

<Appendix 1.>

A-1

IM 12J5B3-01E

Appendix 1. How to Remove the DO30G Membrane Protective

Cap

CAUTION

A black rubber cap is attached to the tip of the DO30G to protect the membrane.
Pulling off the cap forcefully causes a negative pressure inside the protective cap, because the

protective cap sticks to the DO30G, and may stretch the membrane or damage it.
Furthermore, removing the cap by twisting it may loosen the membrane assembly and affect the

measurement results.
Please read the following notes and removal procedure in advance in order to prevent damage

and loosening of the membrane assembly from occurring.

3rd Edition : May 28,2010

1.

Prohibited actions

Prohibited action 1

No good

Figure 1 Do not pull off the cap forcefully

No good

Prohibited action 2

Figure 2 Do not pull off the cap twisting the cap

counterclockwise.

2.

How to Remove Protective Cap

When taking out of the package, a black rubber cap is attached the DO30G shown below.

DO30G

Protective cap

Figure 3 Condition after taking out of the package