2 parts that should be kept on hand – Yokogawa DO30G Dissolved Oxygen Sensor User Manual

Page 22

Advertising
background image

<4. Maintenance>

4-5

IM 12J5B3-01E

Table 4.1 Items Contained in the DOX8A Maintenance Parts Kit

Item

Part No.

Qty

Remarks

Electrolyte

K9171DN 50 ml

Must be replaced periodically, either once every 6 to 8

months or if a span adjustment is not possible. Eight to 9

ml are consumed with each replacement.

Permeable-membrane

assembly / O-ring

K9171HM 3 sets

Normally, these parts must be replaced at the same time

that the sensor is refilled with electrolyte. If the sen-

sor is used in a slurry solution, these parts may require

replacement more often due to abrasion or damage.

Reagent for preparing calibra-

tion solution (for zero adjust-

ment)

sodium sulfite (Na

2

SO

3

)

L9920BR 500 g

Calibration must be carried out at a predetermined

interval (the frequency of calibration varies depending

on the operating conditions). Calibration must also be

performed when, for example, the electrolyte is replaced.

Approximately 15 grams of the reagent is needed to pre-

pare 1 liter of the calibration solution for zero adjustment.

5-ml syringe (without needle)

Needle

L9827NH

L9827NG

1

1

Used when replacing the electrolyte.

200-ml beaker

L9825AF

1

Used when calibration is carried out or when the electro-

lyte is replaced.

Abrasive for silver electrode

(alumina)

K9088PE 30 g

When the permeable-membrane assembly is replaced,

the silver electrode must be polished until it becomes

glossy. Only an extremely marginal amount (approx. 0.5

mg) of the abrasive is consumed with each polishing.

4.3.2

Parts That Should Be Kept on Hand

Even parts with a long service life are not completely free from possible failure. It is recommend-

ed that users have on hand spare parts for those parts which, if they become defective, will make

it impossible to continue measurement.
[Membrane Assembly Locknut (part number: K9171CH)]
It is recommended that users have spare parts for the locknut on hand in preparation for possible

failure.
[Spare Sensor Assembly]
If the sensor assembly remains unused over a prolonged period, the electrolyte may change in

quality or some anomalies may occur in the permeable membrane. Consequently, these ma-

terials must be replaced with new ones before you can use the sensor. Although it is desirable

that users always try to avoid storing the sensor assembly for long periods, the manufacturer

recommends that users keep a spare part on hand for the sensor assembly in anticipation of an

unexpected failure. Do not store the sensor with the electrolyte removed. Otherwise, any residual

electrolyte may become concentrated, harming the lead electrode or other mechanisms. If the

sensor must be stored with the electrolyte removed, completely clean the inside of the sensor

with clean water (repeat injecting and draining the water several times) before it is stored.

Note: If the sensor is stored without the electrolyte, repeat injecting and draining the electrolyte several times before you reuse it.

4th Edition : Apr 05,2011

1

2

3

4

Advertising