<1. General>

1-2

IM 12J5B3-01E

1.2

Specifications

1.2.1

Standard Specifications

Object of Measurement:

Concentration of oxygen dissolved in solution (water)

Principle of Measurement:

Galvanic-cell method

Measurement Range:

0 to 20 ppm, 0 to 20 mg/l, or 0 to 100% (saturation)

Note: The measurement range must be entered through the dissolved oxygen converter.

Resistance Temperture Detector (RTD) :

Pt 1000

Construction:

Sensor unit : Rigid PVC, SUS304

O-ring: Nitrile rubber

Permeable membrane: Fluoro-resin [FEP]

Cable shielding : Heat-resistant flexible PVC

Outlet of cable : Polycarbonate resin

Sensor Cable:

Length : 3 m, 5 m, 10 m, 15 m, or 20 m

(specified when ordering)

Crimping terminal configuration : Pin-type, M4 Ring-type or

Fork-type (specified when ordering )

Weight:

Approx. 0.3 kg + 0.12 × N kg N: cable length

(e.g. for cable length of 5 m, approx. 0.9 kg)

Temperature of Solution Under Measurement (SUM) :

max. 0 to 40 °C

Flow rate of SUM:

No less than 20 cm/s (recommended to prevent the sensor

from trapping bubbles that are the source of error in

measurements)

Pressure of SUM:

Narmally : 0 to 30 kPa

Maximum applicable : 100 kPa

[Characteristics]
Repeatability:

0.1 mg/l or 3% FS, whichever is greater (including sensor)

Temperature Compensation Error:

Within ±3% FS

(including sensor) for a ±5 °C change in the range of 0 to 40 °C

Response Time :

Within 2 minutes (90% response)

(including sensor)

4th Edition : Apr 05,2011