When using a stick or the like – Yokogawa DO30G Dissolved Oxygen Sensor User Manual

Page 25

Advertising
background image

<Appendix 1.>

A-2

IM 12J5B3-01E

When using a stick or the like.

(1) As shown in the figure 4, insert a

stick or the like between the DO30G

and protective cap. Be careful not to

damage the DO30G main body.

Protective

cap

Stick or the like

with a diameter of

0.1 to 0.2 mm

Figure 4 Insert a stick or the like between the

DO30G

(2) Push it in so that enough air space

is created between the DO30G and

protective cap.

Push it beneath the protective cap

and up to the tip of the sensor.

Figure 5 Push a stick or the like beneath the

protective cap to create air space.

DO30G tip

Protective cap

Maintain the space.

Figure 6 Pull off the protective cap while

maintaining the air space.

When an appropriate stick or the

like is not available.

(1) Hold the protective cap while holding the

DO30G as shown in the figure 7.

Figure 7 Hold the DO30G and protective cap.

(2) Remove the protective cap by turning it

clockwise slowly.

Turn it clockwise slowly.

Figure 8 Remove the protective cap by turning it

clockwise slowly

3rd Edition : May 28,2010