<4. Maintenance>

4-1

IM 12J5B3-01E

4.

Maintenance

This section describes the inspection and maintenance procedure applicable to just the dis-

solved oxygen sensor alone. For comprehensive inspection and tuning of the measuring sys-

tem, such as calibration, see the DO402G Dissolved Oxygen Converter instruction manual (IM

12J05D02-01E), the DO202/FLXA21 2-wire Transmitter instruction manual (IM 12J05C01-01E/

IM 12A01A02-01E).

4.1

Routine Inspection

4.1.1

Inspecting the Permeable Membrane (Degree of

Contamination)

An inspection (visual) of the permeable membrane must be made whenever the sensor is cali-

brated. Any dirt on the membrane will adversely affect your measurements. If the dirt is spread

over the membrane, dip the membrane into clean water and gently wipe the dirt off with a soft

cloth or lint-free paper. If the dirt will not come off easily, apply a neutral detergent to the mem-

brane so it will be easier to remove the dirt. Clean the membrane with clean water. Finally, make

sure the permeable-membrane assembly locknut is not loose.

CAUTION

Exercise care to avoid wrinkling or damaging the permeable membrane during inspection.
Wrinkling and other damage prevent normal membrane operation.

4.1.2

Checking of Membrane Assembly Locknut

At every calibration, check that the membrane assembly locknut is not loose. If it is, retighten it.

In addition, if a phenomenon which is considered to be due to a change in the electrolyte qual-

ity (e.g., showing extremely low measured values) is observed after the locknut is retightened,

replace the electrolyte according to the directions in Subsection 4.2.1.