<4. Maintenance>

4-4

IM 12J5B3-01E

4.2.2

Checking of RTD in the Sensor

The resistance temperature detector (RTD) incorporated in the DO30G sensor is a Pt 1000 Ω

RTD. This Pt 1000 Ω RTD indicates a resistance value of about 1097 Ω at 25 °C. Check the RTD

by measuring the resistance value between conductors 11 (T1) and 12 (T2). If the resistance

value differs greatly from the above value, the sensor must be replaced.

4.2.3

Judgment of Lead Electrode Fault

If no failure is found in the checks in Subsections 4.2.1 and 4.2.2, the cause of the problem may

be the end of the service life of the lead electrode or a loose connection of the conductors. In

either case, the sensor must be replaced.
[Service Life of Lead Electrode]
The lead electrode mounted in the sensor body gradually erodes due to battery action. When

span adjustment becomes impossible for a sensor which has been used in succession over a

long period even if the electrolyte/membrane assembly has been replaced, it shows that the lead

electrode has reached the end of its service life. The service life of the lead electrode varies with

the temperature of the SUM or the amount of dissolved oxygen. Under general operating condi-

tions, it should be approximately three to four years.

4.3

Expendables, Spare Parts and

Maintenance Parts

This section presents the general information on spare parts that the manufacturer recommends

you to provide and on materials that are used in maintenance and service. When procuring these

parts or materials, also refer to the Customer Maintenance Parts List (CMPL) at the end of this

manual.

4.3.1

DOX8A Maintenance Parts (Expendables) Kit

The DOX8A Maintenance Parts Kit contains, as a set, spare parts for the expendables that need

to be replaced within one year after the start of use of the DO30G Dissolved Oxygen Sensor, the

materials that are used during replacement, and a reagent that is used to prepare the calibration

solution for zero adjustment.

Note: There are two different DOX8A Maintenance Parts Kit available: style A and style B.

Be sure to use a style B membrane assembly (membrane thickness: 50 μm) for a EXA

DO

dissolved oxygen metering system.