Starting Up

KLEIBER 740

15

6 Starting

Up

6.1

Installation Site Requirements

Climatic conditions

The following requirements must be fulfilled with respect to the climatic conditions at the
place of use of the pyrometer:

temperature range:

0 °C to +40 °C (up to +80°C with connected cooling)

maximum air humidity:

60 % at room temperature

Atmosphere

free of dust, corrosive vapours, and fume gases

Requirements at the place of use

Take into account the following requirements at the place of use of the pyrometer:

Take care with the choice of the place of use and take into account the ergonomic and
the legal guidelines for industrial safety in order to ensure safe operation of the
pyrometer.

Set up the pyrometer on a firm, stable base. The base must be free from concussion
and vibration.
Note:
We recommend using the rail mounting plate available as an accessory for the
attachment of the pyrometer as well as a clamping attachment (see page 22).

Do not expose the system to direct sunlight or radiation from heating (e.g. radiators).

Do not set the pyrometer near strong electromagnetic fields or near electromagnetic
interference sources.

Make sure that the pilot light falls unhindered on the object being measured and that
there is nothing in the path of the rays.

CAUTION

Measurement errors and damage to the pyrometer through
– ambient temperatures too high ,
- strong contamination of the optics due to dust, smoke, steam or other causes

air pollution,

- electromagnetic interference sources

You must take into account the following climatic conditions and the requirements of the
place of use of the pyrometer!