Starting Up

KLEIBER 740

17










Fig. 9

Pyrometer wire connection with configuration

Connector A

Meaning

Pin 1

Supply voltage +24 V (DC or AC)

Pin 2

Supply voltage 0 V (DC or AC)

Pin 3

ground output

Pin 4

Analogue output [0 (4) … 20] mA

Pin 5

empty

Connector B

Meaning

1 - white

Supply voltage +24 V (DC or AC)

2 - brown

Supply voltage 0 V (DC or AC)

3 - green

ground output

4 - yellow

Analogue output [0 (4) … 20] mA

5 – green-yellow

shield

6.2.2 Align

pyrometer

The pyrometer is equipped with an LED pilot light for accurate alignment of the sensor with
the object to be measured. Align the pyrometer with the object to be measured as follows:

1. Switch on the supply voltage to the pyrometer.

2. Allow a starting time of approx. 5 minutes for thermo-stabilization. Than the pyrometer

has stabilised and is ready to work with the given accuracy.

3. Switch on the pilot light by pressing the push button at the rear side. Make sure that

there is nothing in the path of rays.
Note: In order to avoid measuring errors, the area of the pilot light must not be larger
than the object to be measured.

4. If you have a pyrometer with vario optic, adjust the necessary measuring distance (see

page 18).
Note: The macro-optics cannot be adjusted.

The pyrometer is thus aligned and ready for the temperature measurement.