Technical Data

KLEIBER 740

5

Storage temperature

[-20 … +70] °C

Power supply

24 V DC, 0.2 A or 24 V AC, 0.2 A

Degree of protection:

IP 54 according to DIN 40 050

Cooling
− connection

− flow rate
− pressure

− coolant temperature


tube screw connector ¼ "
at least 1 l/min
maximum 6 bar
[10 … 35] °C

Test base

EN 55 011 : 1998, limit class A

CE marking

according to EU regulations

Vario optic

a*

1

in mm

450

600

800

1000 1200 1400 1600 1800 2000 2500 3000

M*

2

in mm

2.5

4.0

6.0

8.0

10.5

11.5

13.0

14.0

15.0

20.0

28.0

*

1

Distance from optics front edge

*

2

Measuring field diameter