Electrical Installation 25

4.2.4 Powering with 115 Volts AC

Note: Voltage/fusing are factory set. Use a wire gauge appropriate to service
installation. See Appendix A for Power and Fuse Requirements and Appendix E

Engineering Drawing 613-816, Service Connections.

To connect 115 VAC to Terminal TB1:

AC Hot to terminal labeled L1/HOT

AC Neutral to terminal labeled N/NEUT

AC Ground to Safety Ground stud.

4.2.5 Powering with 220/230 Volts AC

Note: Voltage/fusing are factory set. Use a wire gauge appropriate to service
installation. See Appendix A for Power and Fuse Requirements and Appendix E
Engineering Drawing 613-816, Service Connections.

North American Connection, as follows to TB1:

AC Hot1 to terminal labeled L1/HOT

AC Hot2 to terminal labeled N/NEUT

AC Ground to Safety Ground stud

European Connection, as follows to TB1:

AC L1 to terminal labeled L1/HOT

AC N to terminal labeled N/NEUT

AC Ground to Safety Ground stud

4.2.6 Powering with 24 Volts DC

Note: Voltage/fusing are factory set. Use a wire gauge appropriate to service
installation. See Appendix A for Power and Fuse Requirements and Appendix E

Engineering Drawing 613-816, Service Connections.

To connect 24VDC to Terminal TB1:

+24Vdc to terminal labeled 24VDC

24V Return to terminal labeled RTN

Ground to Safety Ground stud