2 initial control settings, 3 delay setting, 4 system gain/climate compensation – LumaSense Technologies M8100SM-EXP User Manual

Page 32: 5 oscillation on the analog meter

Advertising
background image

E

2

T QUASAR Flare Monitors Operation Manual

Operating the M8100PM-EXP 32

5.1.2 Initial Control Settings

(Item numbers refer to and Figure 9: Opto-Electronics Module)

1. Adjust all Pots to full counterclockwise (Items: 5, 9, 10).
2. Set Climatic Compensation Selector Switch (Item 4) at 100X.
3. Move the Gain/Setpoint Switch (Item 6) to Setpoint position.
4. Verify the Setpoint Adjust Pot (Item 10) is set for 0.6 on analog meter (Item11).
5. Move the Gain/Setpoint Switch (Item 6) to Gain position after step 4 is accomplished.

Note: When first powered up, the instrument will show a Green LED on even when
no signal is present on meter. This occurs while the Delay circuit is charging and will
switch to Red LED (flame lost status) on, in approximately 2 seconds with the Delay pot

adjusted fully counterclockwise.

5.1.3 Delay Setting

The Delay setting postpones the Alarm condition once the set-point value is exceeded. This

can prevent false alarms from varying conditions or wind.

Once the instrument is aimed at the pilot flame and the Three-Way Switch is set to 100X,

adjust the Delay Pot (item 9) to achieve the desired delay response.

To verify settings, simply block the sight path by covering the Objective lens (item 8) and

time the delay for the red LED to turn on.

Note: Full clockwise for maximum delay (approximately 2.0 minutes).

Full counterclockwise for minimum delay (approximately 2 seconds).

5.1.4 System Gain/Climate Compensation

Optimize these two gain settings by performing the following
(See Figure 9, Opto-Electronics module):

For System Gain, adjust item 5

For Climate Compensation, switch item 4

The purpose is to maximize the System Gain without observing excessive oscillation (noise)
on the analog meter.

5.1.5 Oscillation on the Analog Meter

Some minor fluctuations on the analog meter are acceptable as long as they do not exceed +/-.5
meter units.
Red/Green LED Indicators:
When the analog meter reading falls below the setpoint (0.6 on meter), the red LED “Flame Out
Alarm” will light after the delay has timed out. The green LED will go out.

Advertising
This manual is related to the following products: