Appendix a: specifications and parameter settings – LumaSense Technologies M8100SM-EXP User Manual

Page 55

Advertising
background image

E

2

T QUASAR Flare Monitors Operation Manual

Appendix A: Specifications and Parameter

Settings 55

Appendix A: Specifications and Parameter

Settings

Temperature Range

Designed for Flares/Pilots of any intensity

Working Distance

0 - 1320 ft. (0 - 400m)

Response Time

100 mSeconds

Outputs

Red and Green status lights, mA and set point relay contacts

Relay Contacts

120Vac 0.5A , 24Vdc 1A , Resistive

Sensitivity

Adjustment

0 - 1 VDC

Focusing Range

20 in. (51cm) to infinity

Sighting

Visual aiming

Target Size

(Field of View)

S (dia. of circle) = D (Distance)/37.5, 60, 75, 150, or 300

Power Ratings

115

± 5%, 50/60 Hz (standard)

220

± 5%, 50/60 Hz (optional)

24V

+/-20%, (optional)

115 Watts

Fusing

ELECTRONICS:
115 .250A,’T’ Time-Lag 5x20 Little fuse type 218.250

220 .125A,’T’ Time-Lag 5x20 Little fuse type 218.125
24V

1A,’T’ Time-Lag 5x20 Little fuse type 218.001

6 882 480 Fuse, 1 A (24 VDC), (P/N 15825-1000)

6 882 490 Fuse, 4 A (24 VDC), (P/N 19154-4)

6 882 500 Fuse, 1 A (115 VAC), (P/N 19154-1)
6 882 510 Fuse, 1/4 A (115 VAC), (P/N 19534-0250)

6 882 520 Fuse, 0.5 A (230 VAC), (P/N 19154-0.5)

6 882 530 Fuse, 125 mA (230 VAC), (P/N 19534-0125)

HEATER :

115

1A, Fast TR5, Wickman 370-1100

220 .5A, Fast TR5, Wickman 370-0500
24V

4A, Fast TR5, Wickman 370-1400

Ambient Temperature
Limits

A.

40 to 120 °F (4 to 50 °C) with no cooling

B.

40 to 200 °F (-40 to 80 °C) with cooling base, 10 GPH water flow

at 60

O

F (15 ° C).

C.

40 to 175 °F (-40 to 80 ° C) with vortex air cooler (optional) with

100 psig source.

D. Both cooling methods can accommodate higher ambient

temperature by increasing flow rates. Consult factory.

Advertising
This manual is related to the following products: