as it rotates. The output may produce a non-rising low reading even when looking at high

levels. Inspect the wheel and motor for physical damage. If power is applied and the wheel will
not rotate, check to see if the wheel is hanging up on any obstruction.

Caution: DO NOT attempt to bend the wheel or motor bracket. Instead, loosen the
bracket mounting screws and re-adjust it slightly until free movement is obtained.

8.1.6 Broken Wire Harness

Caution: The disassembly of the Electro-Optical Package will result in the voiding of
the Limited Warranty.

If the instrument appears to be working correctly but the recording or control attachments are

not receiving the proper signals, the problem may be due to a broken wire in the
interconnecting cables. Using field mA and mV calibrators or an ohm meter, check for continuity

of all wiring.

1.

Chassis

Lower

2.

Front

End

PCB

3.

Power

Supply

PCB

4. MOVs

5.

Lens

Screw

6.

Chassis

Upper

7.

Chassis

Assembly

Screw

8.

Mounting

Assembly

Screw

9.

Mains

Fuse

(s)

10.

Heater

Fuse

(s)

Figure 15: Electronics Assembly