Function Generator Calibration
Background
The function generator module can also be
calibrated from the System menu. The calibration
function is only available for firmware version
V1.16 or later.
Connection
Connect the GEN 1 or GEN 2 output (depending
on which module slot the function generator is
installed in) to CH1 using a BNC cable.
CH1
GEN 1
or
GEN 2
Note
If two function generator modules are installed,
the system will automatically choose to calibrate
the function generator in module slot 1. If you
wish to calibrate the second function generator,
first select it from the Option menu (Option key >
Function Gen 2).
Panel Operation
1. Press the Utility key.
Utility
2. Press System from the bottom
menu.
3. Press More 1 of 2 from the side
menu.
4. Press Self CAL from the side menu.