GW Instek GDS-2000 series Programming manual User Manual
Page 25

GDS-2000 Series Programming Manual
25
*RCL
Recall the setting data from memory which previous saved. The settings of
oscilloscope can be stored in memory of S1~S20. However, if users recall the
stored memory which the settings of RS-232 or GPIB are different with present
settings, the RS-232 (or GPIB) settings will keep with the present situation. The
RS-232 (or GPIB) settings will not to be influenced by the recall setting of RS-232
(or GPIB) for this moment.
Syntax
*RCL <NR1>
Arguments:
1~20
Examples
*RCL 1
Recalls the setting data which located at first position of memory address.
*RST (no query form)
Sets all control settings of oscilloscope to their default values but does not purge
stored setting.
Syntax
*RST
*SAV
Saves the setting data to memory.
Syntax
*SAV <NR1>
Arguments
1~20
Examples
*SAV 2
Saves the setting data to the second position of memory queue.