:TEMPlate:POSition:MAX
Sets and query the position of maximum template. This command will not change
the position value which stored in internal memory (W1~W20), unless you save the
template by the command of :TEMPlate:SAVe:MAXimum. That means that if the
oscilloscope is re-opened again without present template storage, the oscilloscope
will recall the previous stored position value from internal memory automatically.
Before running this command, the oscilloscope has to be initialized as “Normal
Template” mode, please add the command of “:TEMPlate:MODe 0” in order to
change the settings.
Syntax
:TEMPlate:POSition:MAX <NR2>
:TEMPlate:POSition:MAX?
Arguments
<NR2> is the desired template position. The range is from -12.0 ~ +12.0 div.
Returns
<NR2>
Note A: 1 division = 25 dots
Note B: Before input any template editing commands, the oscilloscope has to be initialized,
please add the command of “:GONogo:FUNCtion 1” in order to initial the oscilloscope.