2 calibration procedure – zero calibration – KANOMAX 2211 Handheld IAQ Monitor User Manual

Page 44

Advertising
background image

39

8.2 Calibration Procedure – ZERO Calibration

Display

Procedure

Connect ZERO gas tank and IAQ Monitor.

Press


Use

to select [7.CALIBRATION]

Press

.

Use

to select sensor (1.CO2 or 2.CO).

Press

. (CO2 is selected in the example.)

Use

to select [ZERO].

Press

.



Use

to select [1.ZERO], press

.



……

Currently calibrating sensor (CO or CO2)

……

The value before calibration