Specification – KANOMAX 2211 Handheld IAQ Monitor User Manual

Page 49

Advertising
background image

44

9. Specification

Product

IAQ Monitor

Model

2211

Object

Clean air flow

Method

Electrochemical

Range

0.0-50.0 ppm(0.0~500 ppm*)

Resolution

0.0-99.9ppm:0.1ppm,100-500ppm:1ppm

Accuracy

±

3% of displayed value or ±3ppm; whichever that is larger (@20℃)

Compensation: Temp

±

0.125 %FS/℃ (within ‐20-40℃: standard is 20℃)

Compensation:
Air Pressure

±

0.02 %FS/hPa (within 700-1200hPa: standard is 1013hPa)

CO

Response Time

Approx. 60sec. (90%responsive, with calibration cover)

Method

NDIR (Non-Distributed Infrared)

Range

0~5000 ppm

Resolution

1ppm

Accuracy

±

3% of displayed value or ±50ppm; whichever that is larger (@20℃)

Compensation: Temp

±

0.34%FS/℃ (within ‐20-40℃: standard is 20℃)

Compensation:
Air Pressure

±

0.02%FS/hPa (within 700~1200hPa: standard is 1013hPa)

CO2

Response Time

Approx. 45sec. (90%responsive, calibration cover)

Method

Platinum Temperature Measuring Resistive Element

Range

0.0~60.0

°C(‐20.0~60.0

Resolution

0.1

°C

Accuracy

±

0.5

°C

Temp.

Response Time

Less than 60sec. (velocity:1m/s, 90% responsive)

Method

Electrostatic Capacity

Range

2.0~98.0 %RH

Resolution

0.1 %RH

Accuracy

2~80%RH:±2.0%RH,80~98%RH:±3.0%RH

Humidity

Response Time

Approx. 45sec.(90%responsive)

Function

Measured/Max value HOLD, Battery Indicator (5-level), Time/date,
Barometer comp., Unit selection (Temp/DT/WB:

°C or °F, AH:g/m

3

or

lb/ft

3

, OA%:g/kg or lb/lb), Max/Min/Average (interval: 1~999sec., retry:

1~999, Max memory: 1500data), OA%, Gas calibration

Output Function

Digital output: RS-232C (baud rate:4800,9600,19200,38400bps) for
printer and/or PC
Analog Output: DC0~1V (CO, CO2, temperature, or humidity)

Power Supply

6x AA Batteries, AC Adaptor

*

: AC100~240V(50/60Hz)

Battery Life

Approx. 10hours

20℃, alkaline batteries, no RS-232C connection)

Main Body

5~40 ℃ (no condensation)

Probe

20~60 ℃ (no condensation)

Condition

Storage

20~60 ℃ (no condensation)

Weight

Main body: Approx. 400g (including batteries) Probe: Approx. 250g

Standard Kits

Carrying case, Operation manual, AA batteries x6, Calibration cover and
tube, Probe stand, Data processing software for Windows, RS232C
cable

Optional Accesories

Spare probe, analog output, printer, ZERO gas, SPAN gas for CO/CO2,
flow control valve, AC adaptor

Advertising