11. Troubleshooting

11.1 Battery Check

Problem

Possible Cause(s) / Solution(s)

Refer To

(Page No.)

IAQ Monitor will not turn ON

The battery is defective.

Turn OFF the power and replace the batteries.

4,6

Nothing appears on the display

Contrast is not set properly.

Adjust the contrast volume switch.

37

” flashes.

The batteries are low.

Turn OFF the power and replace the batteries.

4,6

11.2 Initial Operation Check

Problem Possible

Cause(s) / Solution(s)

Refer To

(Page No.)

Display is too dark/light

Screen contrast may need adjustment.

Ajust contrast by turning Contrast Adjuster.

37

“NO PROBE” is displayed.

Probe is not connected.

Turn OFF the power and connect Probe.

5

Measurement unit is not
appropriate.

Set appropriate unit of temperature (C,

o

F), absolute

humidity (g/m

3

, lb/ft

3

, and humidity ratio (g/kg, lb/lb).

34

11.3 Check During Measurement

Problem Possible

Cause(s) / Solution(s)

Refer To

(Page No.)

“**.*” is displayed for
measured value.

IAQ Monitor will show “**.*” for the over-the-range
measurement. It must be used within the range to take the
measurement.

44

Probe is not connected property.

Check the connection.

5

“ ‐ ‐ ‐ ‐ ” is displayed for
measured value.

Probe may be damaged.

Contact your local Kanomax Office or service center.

IAQ Monitor is not displaying
the right speed.

Probe sensor may be too close to expiratory air.
Keep a sensor away from expiratory air as much as
possible.

7

Higher temperature is
displayed

.

Theoretically, IAQ Monitorcan not mesure temperature in
no-wind environment.

Gently move probe

8


Humidity reading is lower than
Assman psychrometer.

Assman psychrometer is an intricate instrument and condition
sensitive. Refer to operation manual of the psychrometer.