Technical data – KROHNE OPTIFLEX 2200 C_F EN User Manual

Page 127

Advertising
background image

TECHNICAL DATA

8

127

OPTIFLEX 2200 C/F

www.krohne.com

07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en

Measurement limits (dead zone) in mm and inches

Measurement limits (non-linearity zone) in mm and inches

80 is ε

r

of water; 2.3 is ε

r

of oil

Coaxial probe

Figure 8-10: Measurement limits

1 Device with a coaxial probe
2 Top dead zone:

Top dead zone:

Top dead zone:

Top dead zone: Top part of the probe where measurement is not possible

3 Top non-linearity zone:

Top non-linearity zone:

Top non-linearity zone:

Top non-linearity zone: Top part of the probe with a lower accuracy of ±30 mm / ±1.18¨

4 Bottom non-linearity zone:

Bottom non-linearity zone:

Bottom non-linearity zone:

Bottom non-linearity zone: Bottom part of the probe with a lower accuracy of ±30 mm / ±1.18¨

5 Bottom dead zone:

Bottom dead zone:

Bottom dead zone:

Bottom dead zone: Bottom part of the probe where measurement is not possible

6 Gas (Air)

Gas (Air)

Gas (Air)

Gas (Air)

7 Product

Product

Product

Product

8 L, Probe length

L, Probe length

L, Probe length

L, Probe length

9 Tank Height

Tank Height

Tank Height

Tank Height

10 Minimum distance from the probe to a metallic tank wall:

Minimum distance from the probe to a metallic tank wall:

Minimum distance from the probe to a metallic tank wall:

Minimum distance from the probe to a metallic tank wall: Coaxial probe = 0 mm / 0¨

Dead zone

ε

r

= 80

ε

r

= 2.3

Top 2

Bottom 5

Top 2

Bottom 5

[mm]

[inches]

[mm]

[inches]

[mm]

[inches]

[mm]

[inches]

Coaxial

150

5.91

0

0

200

7.87

20

0.79

Non-linearity

zone

ε

r

= 80

ε

r

= 2.3

Top 3

Bottom 4

Top 3

Bottom 4

[mm]

[inches]

[mm]

[inches]

[mm]

[inches]

[mm]

[inches]

Coaxial

0

0

50

1.97

0

0

150

5.91