Operation – KROHNE OPTIFLEX 2200 C_F EN User Manual

Page 68

Advertising
background image

6

OPERATION

68

OPTIFLEX 2200 C/F

www.krohne.com

07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en

Measurement definitions

Measurement name

Description

Available units

LEVEL

This is a display and an output function option. It

is the height from the bottom of the tank to the

surface of the liquid or solid contents

(Tank height - Distance).

m, cm, mm, in (inches), ft

(feet)

DISTANCE

This is a display and an output function option. It

is the distance from the face of the flange to the

surface of the liquid or solid contents of the

tank.

m, cm, mm, in (inches), ft

(feet)

CONVERSION

This is a display and an output function option. It

gives the volume or mass of the tank contents.

This data is available if you prepare a volume or

mass table in configuration mode. For data on

how to prepare the conversion table, refer to

How to configure the device to measure volume

or mass

on page 85.

kg, t, Ston, Lton, m, cm, mm,

in, ft, m3, L, gal, Imp, ft3, bbl

ULLAGE CONV.

This is a display and an output function option. It

gives the empty volume or remaining mass that

can be put in the tank. This data is available if

you prepare a volume or mass table in

configuration mode. For data on how to prepare

the conversion table, refer to

How to configure

the device to measure volume or mass

on page

85.

kg, t, Ston, Lton, m, cm, mm,

in, ft, m3, L, gal, Imp, ft3, bbl

EPSILON R

The dielectric constant of the contents of the

tank. An electrical property of the liquid or solid

contents of the tank. Also known as ε

r

, DK and

relative permittivity. This gives the strength of

the measurement pulse reflection.

This data is shown if the menu item AUTO Er

CALC (2.5.2) is used to calculate the dielectric

constant value.

OUTPUT I (mA)

The current output of the device.

mA

OUTPUT I (%)

The percentage of the current output.

0% = 4 mA. 100% = 20 mA.

%

Advertising