07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
Default values for menu items 2.4.3 SCALE 4mA and 2.4.4 SCALE 20mA
3. Service menu
4. Master menu
Probe type
SCALE 4mA
SCALE 20mA
[mm]
[inches]
[mm]
[inches]
Ø2 mm / 0.08¨ Single cable
250
9.84
L - 300
1
L - 11.81
1
Ø4 mm / 0.16¨ Single cable
200
7.87
L - 300
1
L - 11.81
1
Single rod
170
6.69
L - 300
1
L - 11.81
1
Coaxial
50
1.97
L - 200
1
L - 7.87
1
Ø4 mm / 0.16¨ double cable
100
3.94
L - 250
1
L - 9.84
1
Ø8 mm / 0.31¨ double rod
110
4.33
L - 270
1
L - 10.63
1
1 L, Probe length
Menu
No.
Function
Function description
Selection list
Default
3.0.0
SERVICE
Advanced settings. The settings in this
menu are protected with a password.
Only approved personnel can change
the parameters in this menu. For more
data, contact your local sales office.
Menu
No.
Function
Function description
Selection list
Default
4.0.0
MASTER
Factory settings. The settings in this
menu are protected with a password.
Only approved personnel can change
the parameters in this menu. For more
data, contact your local sales office.