ERR 209 Sensor not
compatible
1
The software version of the
sensor is not compatible with the
software version of the signal
converter.
Go to menu 1.1.0 IDENT. in
Configuration mode. Record the
version numbers of the device
software given in menu items
1.1.2, 1.1.3 and 1.1.4. Give this
data to the supplier.
1
Defective wiring.
ERR 210 Empty
2, 4
The level is in the bottom dead
zone. There is a risk that the tank
is empty.
Add some of the product until
the level is above the bottom
dead zone.
INFORMATION!
In 4.0.0 MASTER menu, the type of error shown for error codes 102, 201 and 203 can be changed
from "Error" to "Warning" (the NE 107 status signal changes from "Failure" to "Out of
specification"). For more data, speak to the supplier.
Error
code
Error Message
Status marker
shown
Cause
Corrective action